"I think that the data we produced [on Maurice] shows excellent quality, due to the precise and reliable Maurice instrument."
Professor Hermann Wätzig, Technical University of Braunschweig in Germany
View all citations for the Imaged Capillary Electrophoresis (iCE) platform.
Please visit our Instrument Citation Database to search our collection of scientific articles that feature our instruments.
"I think that the data we produced [on Maurice] shows excellent quality, due to the precise and reliable Maurice instrument."
Professor Hermann Wätzig, Technical University of Braunschweig in Germany
"The short run times and easy sample preparation allowed for method development within a day or two. The versatility of the instrument allows the possibility of offering clients rapid alternatives to traditional ID and charge heterogeneity assays."
Joan Garrison, QC Method Transfer Scientist, Cook Pharmica
Feature | cIEF Specification |
---|---|
Minimum Sample Volume | 50 µL |
Sample Delivery | Vacuum |
Typical Separation Time | 6-10 min (molecule-dependent) |
Detection Capability |
UV Absorbance at 280nm Fluorescence: Ex 280nm, Em 320-450nm |
Typical Voltage | Pre-focusing: 1500 V, focusing: 3000 V |
Sample Injections per Cartridge | 100 |
Maximum Sample Injections per Batch | 100 guaranteed, 200 maximum |
pl Range | 2.85-10.45 |
pl CV | 1% |
CV for Peaks >10% Composition | ≤5% (Intra-batch), ≤6% (Inter-batch) |
pl Resolution | 0.05 pl units (for wide range 3-10 ampholytes) |
Dynamic Range | 2 logs |
Linearity | >0.995 |
Sensitivity (LOD) |
0.7 µg/mL (Native fluorescence) 3.0 µg/mL (Absorbance) (Values based on a monoclonal antibody) |
Sample Tray Options | 96-well plates or 48 vials |
Dimensions | 44cm H x 42 cm W x 61 cm D |
Weight | 46 kg (100 lb) |